The Therapeutic Goods Administration (TGA) requires from suitably qualified and experienced providers a HPLC System to support the physico/chemical analysis of biological medicines

 

Biochemistry section requires a configurable and customisable HPLC system for analytical testing and evaluating of biological medicines, proteins and polysaccharides with focus on the quality attributes.

 

The proposed system requires:

 

A pump

 

capable of delivering precise quaternary gradients and including an integrated solvent degasser

 

An auto-sampler

 

with a minimum injection volume range of 0.1 – 100 µL.

 

with demonstrated low carry over between injections.

 

with a sample compartment thermostatted from 4 – 40 °C.

 

that is compatible with standard 2 mL HPLC vials

 

A column compartment

 

With capacity for at least two 300 x 7.8 mm HPLC columns

 

Capable of temperature control from ambient to 80°C

 

Details for options such as active preheaters, column switching valves and cooling of column compartment should be provided. 

 

Detectors

 

1) Options for a photodiode array detector (for each HPLC system) must be included. The detector should have:

 

a minimum wavelength range of 190 - 600 nm

 

a minimum wavelength accuracy of ±1 nm

 

a minimum data sampling rate of 80 Hz (that is the max)            

 

2) Options for one additional UV-Fluorescence detector must be included.

 

The detector should have:

 

excitation range 200-1200 nm and emission range 200-1200 nm

 

Minimum 74 Hz data acquisition rate

 

Long life lamp (>4000 hr) 

 

The systems must be integrated into an existing Waters Empower CDS environment. Software to perform Preventative maintenance and instrument diagnostics  must be provided with the instrument (or details of how to perform PM in a Waters Empower CDS environment).

:

 

As per clause 11 of the RFT

 

              The Tenderer must not have had any judicial decisions against it (excluding decisions under appeal) relating to employee entitlements and have not paid the claim.

 

              The Tenderer, its personnel, and any Subcontractors proposed in the Tender must not, at the Closing Time, be listed as terrorists under section 15 of the Charter of the United Nations Act 1945 (Cth).  

              The Tenderer (and any Subcontractor proposed in its Tender) must not be named in the Consolidated list referred to in Regulation 40 the Charter of United Nations (Dealing with Assets) Regulations 2008 (Cth).

:

31 May 2018

:

AusTender Website (www.tenders.gov.au)

Contact Details

Contact Officer

: 02

: